Archive for September, 2013

EC1 – CONDUCTIVITY / TDS / SALINITY / TEMPERATURE METER

Posted by admin
EC1-Square-for-web

KEY FEATURES

    • Reads Conductivity, TDS, Salinity & Temperature.
    • Temperature range; 0 to 90 °C.
    • Conductivity range; 0 to 2000 µS, 2.00 to 20.00 mS.
    • TDS range; 0 to 1300 ppm, 1.30 to 13.00 ppt.
    • Salinity Range; 0 to 1000 ppm.
    • Floats in water.
    • Auto shut off.
    • Data hold.
    • Only weighs 74 g.

EC1 CONDUCTIVITY/TDS/TEMPERATURE METER

The EC1 is a simple to use, accurate, reliable hand held meter which has been designed to handle the tough environment of field work. The instrument provides the user with readings for conductivity/ TDS and temperature providing maximum simplicity.

CALIBRATION SOLUTIONS

Each meter is supplied with a certificate of conformity which is valid for 12 months.

All types of meters can be sent to our laboratory for an annual calibration which is valid for 12 months.

ORDERING INFORMATION

ORDER CODE
DESCRIPTION
EC1
Conductivity/TDS/Temperature Meter
EC1P
Replacement EC1 probe

CALIBRATION SOLUTIONS

ORDER CODE
DESCRIPTION
VOLUME
SDT109
1413 µS Conductivity Solution
65ml

TECHNICAL SPECIFICATION

PARAMETER
SPECIFICATION
CONDUCTIVITY
Range:
0 to 2000 µS / 2.00 to 20.00 mS
Resolution:
1 µS / 0.01 mS
Accuracy:
1 µS / 0.01 mS
TOTAL DISSOLVED SOLIDS (TDS)
Range:
0 to 1300 ppm / 1.30 to 13.00 ppt
Resolution:
1 ppm / 0.01 ppt
Accuracy:
1 ppm / 0.01 ppt
SALINITY
Range:
0 to 1000 ppm / 1.00 to 12.00 ppt
Resolution:
1 ppm / 0.01 ppt
Accuracy:
1 ppm / 0.01 ppt
TEMPERATURE
Range:
0 to 90 °C
Resolution:
0.1 °C
Accuracy:
±0.2 °C
METER
Dimensions:
170 x 40 x 40 mm (L x W x H)
Weight:
74 g

 

pH1 – pH/TEMPERATURE METER

Posted by admin
PH1-square-for-web

KEY FEATURES

    • Reads pH & Temperature.
    • Temperature range; 0 to 90 °C.
    • Accuracy; ±0.2 °C.
    • pH range; -2 to 16.00.
    • Accuracy; ±0.01 + 1 digit
    • Unique robust electrode almost unbreakable.
    • Floats in water.
    • Auto shut off.
    • Data hold.
    • Only weighs 85 g.

pH1 – pH/TEMPERATURE METER

The DTK range of pocket waterproof testers will float in water and give you the flexibility to monitor water conditions in rugged conditions.

The pH1 uses the unique flat electrode which means it is very robust. It displays pH and temperature simultaneously along with the battery life. It will read from -2 to 16.00 at a resolution of 0.01 pH and has ATC. The unit only weighs 85 grams and has an auto shut off feature. The pH1 can also be used with the RE1P probe to allow the meter to measure redox in mV from -1000 to 1000 mV.

CALIBRATION SOLUTIONS

Each meter is supplied with a certificate of conformity which is valid for 12 months.

All types of meters can be sent to our laboratory for an annual calibration which is valid for 12 months.

ORDERING INFORMATION

ORDER CODE
DESCRIPTION
pH1
pH/Temperature Meter
pH1P
Replacement pH1 probe
pH1R
pH meter with redox probe reading mV
RE1P
Replacement redox probe

CALIBRATION SOLUTIONS

ORDER CODE
DESCRIPTION
VOLUME
SDT119
pH 4 Buffer Solution
65 ml
SDT118
pH 7 Buffer Solution
65 ml
SDT161
pH 10 Buffer Solution
65 ml

TECHNICAL SPECIFICATION

PARAMETER
SPECIFICATION
pH
Range:
-2 to 16.00 pH
Resolution:
0.01 pH
Accuracy:
± 0.01 + 1 digit
TEMPERATURE
Range:
0 to 90.0 °C
Accuracy:
±0.2 °C
METER
Dimensions:
170 x 40 x 40 mm (L x W x H)
Weight:
85 g

 

UNIVERSAL RAPID LEGIONELLA TEST KIT

Posted by admin

Universal Rapid Legionella1000x1000

KEY FEATURES

  • Ideal for mobile treatment engineers.
  • Compact & portable design.
  • Rapid detection of Legionella Bacteria.
  • Results achieved in 30-35 minutes.
  • Detects Legionella pneumophila serogroup 1 bacteria.
  • Easy to follow test procedure.
  • Everything included to perform 5 tests.

UNIVERSAL RAPID LEGIONELLA TEST KIT

The Universal Rapid Legionella Test kit is a breakthrough in technology from Hyrdosense for the rapid detection of legionella bacteria in industrial water. Test results can be obtained in minutes instead of days.

The industrial legionella kits provides the user:

  • Fast results.
  • High sensitivity.
  • In-line sampling.

 

The test kit contains all the items required to perform 5 tests – including filters, recovery buffer, collection vessel and sample point.

LEGIONELLA CONTROL

This Legionella field test kit uses immunochromatographic assay, to detect the presence of cell surface antigens from Legionella bacteria in a sample.

The presence of antigen causes the ‘Test Line’ to turn red in colour. A ‘Control Line’ is included which should always turn red if the test has been performed correctly.

These tests are optimised for use with industrial water systems, cooling towers and whirlpool spas.

The filtration step needs no pumping, just connect the sample point to the water system and use the system pressure to drive the filtration.

The test follows a simple procedure and the results easily obtained within 30 to 35 minutes. Once the result is obtained the treatment engineer can asses the water condition and complete any treatment necessary to ensure there is no risk of out breaks and subsequent litigation.

TEST KIT & INSTRUMENTS

ORDER CODE DESCRIPTION Qty
LTK0051
Universal Rapid Legionella Test Kit
KIT
EDT078L
50ml Luer Lock Syringe
5
EDT083
Plastic Wide Neck Bottle 500ml
1
LTK0070
Legionella Test Cell
5
LTK0080
Legionella Filter
5
LTK0090
Legionella Syringe
5

REPLACEMENTS

ORDER CODE DESCRIPTION
PACK SIZE
LTK0051F
Universal Rapid Legionella Test Kit Refill
ea
EDT078L
50ml Luer Lock Syringe
ea
LTK0070
Legionella Test Cell
ea
LTK0080
Legionella Filter
ea
LTK0090
Legionella Syringe
ea

 

PRODUCT DETAILS

Bacteria detected Legionella pneumophila serogroup 1
Detection method Lateral flow immunochromatographic assay
Lower limit of detection Up to 100 CFU/L (0.1 CFU/mL)
Specificity Tested against a wide range of waterborne bacteria, including
Acinetobacter calcoaceticus, Aeromonas hydrophila subsp.
hydrophila, Bacillus subtilis, Burkholderia cepacia, Citrobacter freudii,
Citrobacter koseri, Escherichia coli, Enterobacter cloacae, Klebsiella oxytoca, Pseudomonas aeruginosa, Pseudomonas fluorescens,
Pseudomonas putida, Pseudomonas stutzeri, Ralstonia pickettii,
Raoultella terrigena, Streptococcus pyrogenes, Yersinia ruckeri. No
cross reactions observed at concentrations > 109 CFU/L.Staphylococcus aureus has been observed to cross react with the test at concentrations > 108 CFU/L.Legionella pneumophila serogroups 4 & 7 cross react with the test at concentrations >109 CFU/sample. The test does not react with any other L. pneumophila serogroups or L. species tested.
Test volume 100 μL
Suggested sample volume 250 mL
Sample parameters Sample should filter in less than 10 minutes.
Biocides The product has been tested with a range of common biocides and
iodispersants used in Legionella control, at their normal maximum perating concentration. Active ingredients tested include:
lutaraldehyde, didecyl-dimethyl-ammonium chloride, isolthazoles, dibromonitrilopropionamide, ionic and non-ionic surfactants.
The test gives a false positive if used with polymeric biguanide or
THPS.
Operator time 5 – 10 minutes
DTK Water
14 Morris Close,
Park Farm Ind Est,
Wellingborough,
NN8 6XF, UK
T: +44 (0)1604 686995
F: +44 (0)1604 686997
E: info@dtkwater.com

DTK Water is a Lovibond® company
Privacy Policy